ELISA for human CCL2/MCP-1
R&D Systems ELISA for human CCL2/MCP-1
Judith Sunzenauer (last update 11.6.2013)
Procedures
Preparation:
- Bring all reagents and samples to room temperature before use.
- Run duplicates.
- Prepare Wash Buffer: Add 20ml of Wash Buffer Concentrate to 480ml a.dest (1:25 Dilution)
- Discard Calibrator Diluent RD5L Concentrate (for urine/cell culture spernate only!!)
- Reconstitute CCL2 Standard: add 5ml Calibrator RD6Q to vial of CXCL12 Standard (concentration of stock: 2000pg/ml). Allow the standard to sit for a minimum of 15min prior to making dilutions.
- Dilute Plasma samples: 1:4 with Calibrator Diluent RD6Q
- Produce the dilution series for CCL2 Standardcurve:
Prepare 8 prelabeled (#1-8) polypropylene tubes and fill each tube with 500µl of Calibrator Diluent RD6Q. Transfer 500µl Standard from the original vial to tube #1, mix thoroughly and transfer 500µl from tube #1 to #2,….. continue until tube #7. Fill tube #8 with Calibrator Diluent RD6Q.
Assay Procedure:
- Add 50µl of Assay Diluent RD1-83 to each well.
- Add 200µl of Standard Dilution #1-8 or sample per well.
- Seal the Plate and incubate 2h at room temperature.
- Aspirate each well.
- Wash 3 times with 400µl Wash Buffer.
- After the last wash, remove any remaining liquid and invert the plate and blot it against clean paper towels.
- Add 200µl of CCL2/MCP-1 Conjugate to each well.
- Seal and incubate for 2h at RT.
Prepare Substrate Solution: mix color Reagents A and B in equal volumes within 15min of use. Protect from light!
- Wash 3 times with 400µl Wash Buffer.
- Add 200µl of Substrate Solution to each well.
- Incubate for 30min at RT and protect from light.
- Add 50µl Stop Solution to each well. Gently tap the plate to ensure mixing.
- Determine the OD at 450nm within 30min. Set wavelength correction to 540nm or 570nm.
Calculation of Results:
- If samples have been diluted, the concentration read from the standard curve must be multiplied by the dilution factor.
- Concentrations of the Standard Dilutions:
Std. # | CCL2 pg/ml |
1 | 1000 |
2 | 500 |
3 | 250 |
4 | 125 |
5 | 62.5 |
6 | 31.2 |
7 | 15.6 |
8 | 0 |
Sensitivity:
Assay | Mean in EDTA Plasma(pg/ml) | Range (pg/ml) |
CCL2 | 153 | 1,7-2.000 |
Handling Photometer + Software:
Make sure the laser is warmed up prior to scanning for 15min
Materials
CCL2 Human Immunoassay |
R&D Systems |
#SCP00 |